General Product Information
Error Terms
Error Term Data
The error term descriptions in this section include the following information:
— a table of error terms
— description and significance of each error term
— measurements affected by each error term
— typical cause of failure for each error term
The same description applies to both the forward (F) and reverse (R) terms.
Data are listed here as a convenience only. Detailed instrument specifications are
in the Data Sheet, viewable online at
http://literature.cdn.keysight.com/litweb/5992-2765EN.pdf
Table 6-1
Frequency Range
300 kHz to < 2 MHz
2 MHz to 1 GHz
> 1 to 2 GHz
> 2 to 4 GHz
> 4 to 6.5 GHz
> 6.5 to 9 GHz
> 9 to 14 GHz
> 14 to 20 GHz
> 20 to 24 GHz
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system
verification. Refer to
92
Error Term Data
Directivity
1. The data in this table are uncorrected system performance. The values apply over an envi-
ronmental temperature range of 25 °C ±5 °C, with less than 1 °C deviation from the cali-
bration temperature.
2. All crosstalk values are typical.
"System Verification" on page
1
Source
Load match
match
9
9
21
19
21
20
21
20
20
15
11
11
9
9
4
6
3
5
Keysight P937xA PXIe Service Guide
.
2
Crosstalk
9
97
21
95
19
123
15
121
11
121
9
119
7
110
6
98
4
82