Keysight P937 A Series Service Manual page 36

Streamline usb vector network analyzers
Table of Contents

Advertisement

Tests, Adjustments, and Troubleshooting
Preliminary Checks
Accessories Used in the Operator's Check
Performing the Operator's Check
1. On the soft front panel, click Utility > System > Service > Verification >
2. In the Operator's Check dialog box (refer to
3. Click Begin.
4. If shorts and opens are not connected to all ports, you will be prompted to
5. The result of the Operator's Check will be shown as a PASS or FAIL to each
34
Male short(s) or open(s) from a 3.5 mm calibration kit.
Operator's Check.
click either Prompt for attachment of Short/Open, to pause at each step in
the process to allow moving the short/open to the appropriate port, or
Shorts/Opens are attached to ALL ports, to run through the test without
stopping. Shorts and opens can be mixed on the test ports.
In most cases, the short/open is not required since the analyzer test port's
female connector acts an "open." However, if the test fails, confirm the
failure with a short/open attached.
connect them as they are needed.
test (refer to
Figure
3-3).
Figure
3-3), under Configure,
Keysight P937xA PXIe Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

P9370aP9372aP9373aP9374aP9375aP9371a