Tests, Adjustments, and Troubleshooting
System Verification
System Verification Procedure
1. If you want printed test outputs, connect a printer to the embedded or
2. Let the analyzer warm up for at least 15 minutes.
3. Insert the verification kit memory stick in a USB port of the embedded or
4. Click Utility > System > Service > System Verification. The System
Figure 3-5 System Verification Dialog Box
5. In the Select Calibration Kit box, click on the calibration kit or electronic
6. Under Outputs, click one of the following options. Refer to
40
remote controller. For the printer, ensure that the correct driver is loaded
and the printer is defined as the default printer. Refer to the network
analyzer's Help system for information on printer setup.
remote controller.
Verification dialog box is displayed, as shown in
calibration module (ECal) being used. The corresponding verification kit to
use is selected for you and displayed in the Compatible Verification Kit(s)
box.
—Print Tabular Data: Prints the verification data in tabular form which
includes measured data and uncertainty limits. For an example, refer
to
Figure 3-7 on page
—Print Graphs: Prints the verification data in graphical form. The
graphical form includes measured data trace, factory supplied data
trace, and uncertainty limits. For an example, refer to
page
44.
Figure
43.
Keysight P937xA PXIe Service Guide
3-5.
Figure
3-5.
Figure 3-8 on