Error Terms
Figure 6-2
Flowgraph of 2-Port Error Terms
Where:

Manual Access to Error Terms

Soft Front Panel Access to Error Terms
Ensure the calibration correction is active by making sure that the check
box for Correction on/OFF has been selected.
90
E = Error term
1st Subscript:
D = Directivity
S = Source Match
R = Reflection Tracking
X = Crosstalk (Isolation)
L = Load Match
T = Transmission Tracking
2nd Subscript:
F = Forward measurement (Port 1 to Port 2)
R = Reverse measurement (Port 2 to Port 1)