General Product Information
Error Terms
Directivity (E
E
DF
the system. The directivity error of the test port is determined by measuring the
S
11
return loss specification than does the uncorrected text port. Therefore, any
power detected from this measurement is assumed to be from directivity error.
The measurements most affected by directivity errors are measurements of low
reflection devices.
Typical Cause of Failure
The calibration kit load is the most common cause of directivity specification
failure.
If the load has been gaged and its performance independently verified, the
analyzer requires repair. Refer to
or Repair" on page
Source Match (E
E
SF
the driven port. They are obtained by measuring the reflection (S
open, and a short that are connected directly to the ports. Source match is a
measure of the match of the coupler, as well as the match between all
components from the source to the output port.
The measurement most affected by source match errors are reflection and
transmission measurements of highly reflective DUTs.
Typical Cause of Failure
The calibration kit open or short is the most common cause of source match
specification failure.
If the open or short performance are independently verified, the analyzer
requires repair. Refer to
Repair" on page
Load Match (E
Load match is a measure of the impedance match of the test port that
terminates the output of a 2-port device. The match of test port cables is
included in this response. Load match error terms are characterized by
measuring the S
calibration procedure.
The measurements most affected by load match errors are all transmission
measurements, and reflection measurements of a low insertion loss, 2-port
device, such as an airline.
Keysight P937xA PXIe Service Guide
and E
DF
DR
and E
are the uncorrected forward and reverse directivity error terms of
DR
and S
reflection of the calibration kit load. The load has a much better
22
22.
and E
SF
and E
are the forward and reverse uncorrected source match terms of
SR
"Shipping Your Network Analyzer for Service or
22.
and E
LF
LR
and S
11
22
)
"Shipping Your Network Analyzer for Service
)
SR
)
responses of a "thru" configuration during the
, S
) of an
11
22
93