Keysight P937 A Series Service Manual page 93

Streamline usb vector network analyzers
Table of Contents

Advertisement

General Product Information
Error Terms
To access the error terms from the soft front panel, perform the following
steps:
1. On the P937xA software's toolbar, click Response > Cal > Cal Sets & Cal
2. In the Cal Sets drop-down list, select the desired cal set or use the System
3. Click Standards to view the raw measurement data of the Standard (ONLY
4. Use the Error Term box drop-down list to select an error term or standard to
5. Select the Enable check box to view the data on the PC display.
6. Compare the displayed measurement trace to the equivalent data that
7. Print numerical data or print a plot of the measurement results.
Keysight P937xA PXIe Service Guide
Kits > Cal Set Viewer.
The Cal Set Viewer tool bar appears directly above the trace window.
Default.
available with Unguided Cal (not ECal or Guided Cal)), or click Error Terms
to view the corrected error term data.
view.
starts in
"Error Term Data" on page
to the uncorrected performance specifications listed in
page
92.
92, or to previously measured data, or
Table 6-1 on
91