Tests, Adjustments, and Troubleshooting
System Verification
For printed output, it is assumed that the printer has been tested and the
Windows driver is installed for the printer that is being used. The system
verification test prints to the printer that has been designated as the
default printer. (On the Windows Desktop display, click on Start > Devices
and Printers to verify the printer setup.
7. To modify the number of ports to be verified or to change the number of
8. Click Run.
9. Follow the instructions for performing a full calibration or recalling an
10.Follow the instructions for performing the system verification, inserting the
If the System Fails the Verification Test
Inspect all connections. DO NOT remove the cable from the network
analyzer test port. This will invalidate the calibration that you performed
earlier.
1. Disconnect and clean the device that failed the verification test.
2. Reconnect the device, making sure that all connections are torqued to the
3. Measure the device again.
4. If the analyzer still fails the test, check the measurement calibration by
5. Refer to
Keysight P937xA PXIe Service Guide
—File Tabular Data: Writes the tabular data to a text file in the
C:\Users\Public\Public Documents\Network Analyzer\SysVer\
directory.
—File Graphs: Saves a screen image in PNG format in the
C:\Users\Public\Public Documents\Network Analyzer\SysVer\
directory.
devices to measure, click on the Configure tab and make the desired
selections.
existing recent calibration.
verification devices as prompted.
proper specifications.
viewing the description of the error terms. For more information, refer to the
TME webpage at http://cal.software.keysight.com.
Figure 3-6 on page 42
for additional troubleshooting steps.