Power Offset Of The Tracking Generator - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
4.8.7

Power Offset of the Tracking Generator

POWER
POWER SWP ON/OFF
SWEEP
START POWER
STOP POWER
The POWER SWEEP softkey opens a submenu for activates or deactivates the
power sweep.
POWER SWP
The POWER SWP ON/OFF softkey activates or deactivates the power sweep. If the
ON/OFF
power sweep is ON the enhancement label TGPWR is shown and the Spectrum
Analyzer is set in zero span mode (span = 0Hz). During the sweep time of the zero
span the power at the internal tracking generator is changed linear from start power
to stop power. The start and stop power values are shown on the right side below
the diagram.
Remote command:
START POWER
The START POWER softkey defines the start power of the power sweep.
The start power can be set between -30 dBm and +0 dBm.
Remote command:
STOP POWER
The STOP POWER softkey defines the stop power of the power sweep.
The start power can be set between -30 dBm and +0 dBm.
The stop value can be smaller than the start value.
Remote command:
Operating Manual 1164.4556.12 - 06
Tracking Generator – Option R&S FSP-B9
:SOUR:POW:MODE SWE
:SOUR:POW:MODE FIX
:SOUR:POW:STAR –20dBm
:SOUR:POW:STOP –10dBm
Instrument Functions
4.225