1Selection Of Detector - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
Example - data section of the file
File contents
Trace 1:;;
x-Unit;Hz;
y-Unit;dBm;
Preamplifier;OFF;
Transducer;OFF;
Values; 625;
10000;-10.3;-15.7
10180;-11.5;-16.9
10360;-12.0;-17.4
...;...;
DECIM SEP
The DECIM SEP softkey selects the decimal separator between '.' (decimal point)
and ',' (comma) with floating-point numerals for the
With the selection of the decimal separator different language versions of evaluation
programs (e.g. Microsoft Excel) can be supported.
Remote command:
COPY TRACE
The COPY TRACE softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory is overwritten and the new con-
tents displayed in view mode.
Remote command:
4.5.8.1
Selection of Detector
The detectors of the R&S FSP are implemented as pure digital devices. The following
detectors are available:
The peak detectors compare the current level value with the maximum or minimum
level from the previously sampled data. When the number of samples defined by the
instrument setting is reached, the samples are combined in the displayed pixels.
4.44
"Peak detectors (MAX PEAK and MIN PEAK)" on page 4.45
"Auto peak detector" on page 4.45
"Sample detector" on page 4.45
"RMS detector" on page 4.45
"Average detector" on page 4.46
"Quasipeak detector" on page 4.46
Description
Selected trace
Unit of x values:
Hz with span > 0; s with span = 0;
dBm/dB with statistics measurements
Unit of y values:
dB*/V/A/Ω depending on the selected unit with y axis LOG or % with
y-axis LIN
Preamplifier state (OFF, if no preamplifier is available)
Transducer state:
OFF:No transducer active (or available)
<trd name>selected transducer
Number of measurement points
Measured values:
<x value>, <y1>, <y2>
<y2> being available only with detector AUTOPEAK and containing
in this case the smallest of the two measured values for a test point.
ASCII FILE EXPORT
FORM:DEXP:DSEP POIN
TRAC:COPY TRACE1,TRACE2
Operating Manual 1164.4556.12 - 06
Instrument Functions
Analyzer Mode
function.