Saving And Recalling Data Sets - File Key; 1Overview - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
Remote command:
OPEN START
The OPEN START MENU softkey opens the windows XP start menu and therefore
MENU
provides an easy access to standard windows functions if a mouse is connect.
POWER
The POWER METER softkey opens a submenu for measurements with power sen-
METER
sors. For details refer to the Software Manual of the application firmware R&S FS-
K9 on the options CD-ROM.
4.6.4
Saving and Recalling Data Sets – FILE Key
4.6.4.1
Overview
The FILE key calls the following functions:
The R&S FSP is capable of internally storing complete instrument settings with
instrument configurations and measurement data in the form of data sets. The
respective data are stored on the internal hard disk or, if selected, on a memory stick
or on a floppy. The hard disk and floppy disk drives have the following names:
It is possible to load data sets stored with another instrument of the same analyzer
family. However, keep the following restrictions in mind:
FILE
SAVE | RECALL
Operating Manual 1164.4556.12 - 06
Storage/loading functions for storing (SAVE) instrument settings such as
instrument configurations (measurement/display settings, etc.) and measurement
results from working memory to permanent storage media, or to load (RECALL)
stored data into working memory.
Functions for management of storage media (FILE MANAGER). Included are
among others functions for listing files, formatting storage media, copying, and
deleting/renaming files.
floppy disk A:, hard disk D:
The frequency range of the loading instrument must include the range of the
instrument used to store the data set.
Example: A R&S FSP40 can load a data set, saved with a R&S FSP26 but a
R&S FSP8 can not load a data set saved with a R&S FSP26. A R&S FSP with
a frequency range of at least 26 GHz is required.
An instrument with an additional option can load a data set saved with an
instrument without this option, but not the other way round.
Example: A FSQ 8 with option R&S FSP-B17 can load a data set, saved with a
R&S FSP8 without this option but not the other way round.
!
SAVE FILE
RECALL FILE
SELECT PATH
SELECT FILE
EDIT FILE NAME
SYST:APPL:SREC ON
Instrument Functions
Basic Settings
4.193

Hide quick links:

Advertisement

Table of Contents