Transmission Measurement; 1Calibration Of Transmission Measurement - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
The permissible setting range is -200 dB to +200 dB in steps of 0.1 dB. Positive off-
sets apply to an amplifier and negative offsets to an attenuator subsequent to the
tracking generator.
The default setting is 0 dB. Offsets <> 0 are marked by the activated enhancement
label LVL.
Remote command:
4.8.2

Transmission Measurement

This measurement will yield the transmission characteristics of a two-port network.
The built-in tracking generator serves as a signal source. It is connected to the input
connector of the DUT. The input of the R&S FSP is fed from the output of the DUT.
Fig. 4.19 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g.
frequency response of connecting cables).
4.8.2.1
Calibration of Transmission Measurement
SOURCE CAL
CAL TRANS
CAL REFL SHORT
CAL REFL OPEN
NORMALIZE
REF VALUE POSITION
REF VALUE
RECALL
SAVE AS TRD FACTOR
The SOURCE CAL softkey opens a submenu comprising of the calibration functions
for the transmission and reflection measurement.
4.214
Tracking Generator – Option R&S FSP-B9
SOUR:POW:OFFS -10dB
Operating Manual 1164.4556.12 - 06
Instrument Functions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents