Model-Dependent Default Settings; Data Output Formats - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
6.23.7

Model-Dependent Default Settings

If the GPIB language is switched over to an 85xx model, the GPIB address will automatically be switched
over to 18 provided that the default address of the R&S FSP (20) is still set. If a different value is set, this
value will be maintained. Upon return to SCPI, this address will remain unchanged.
The following table shows the default settings obtained after a change of the GPIB language and for the
commands IP, KST and RESET:
Model
# of Trace Points
8566A/B
1001
8568A/B
1001
8560E
601
8561E
601
8562E
601
8563E
601
8564E
601
8565E
601
8594E
401
Notes regarding the set stop frequency:
Note regarding the number of test points (trace points):
6.23.8

Data Output Formats

In the case of the SCPI and IEEE488.2 standards, the output formats for numerical data are flexible to a
large extent. The output format for the HP units, by contrast, is accurately defined with respect to the
number of digits. The memory areas for reading instrument data have therefore been adapted accord-
ingly in the remote-control programs for instruments of this series.
Therefore, in response to a query, the R&S FSP returns data of the same structure as that used by the
original instruments; this applies in particular to the number of characters returned.
Two formats are currently supported when trace data is output: Display Units (command O1) and physi-
cal values (command O2, O3 or TDF P). As to the "Display Units" format, the level data of the R&S FSP
is converted to match the value range and the resolution of the 8566/8568 series. Upon transition to the
6.278
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E
Start Freq.
2 GHz
0 Hz
0 Hz
0 Hz
0 Hz
0 Hz
0 Hz
0 Hz
0 Hz
The stop frequency given in the table may be limited to the corresponding
frequency range of the R&S FSUP.
Command LF sets the stop frequency for 8566A/B to a maximum value of 2
GHz.
The number of trace points is switched over only upon transition to the REMOTE
state.
Remote Control – Description of Commands
Stop Freq.
Ref Level
22 GHz
0 dBm
1.5 GHz
0 dBm
2.9 GHz
0 dBm
6.5 GHz
0 dBm
13.2 GHz
0 dBm
26.5 GHz
0 dBm
40 GHz
0 dBm
50 GHz
0 dBm
3 GHz
0 dBm
Operating Manual 1164.4556.12 - 06
Input Coupling
AC
AC
AC
AC
AC
AC
AC
AC
AC