The RECALL softkey restores the instrument setting with which the calibration was
carried out.
This can be useful if the device setting was changed after calibration (e.g. center fre-
quency setting, frequency deviation, reference level, etc.).
The softkey is only available if:
the NETWORK mode has been selected
the memory contains a calibration data set.
Remote command:
SAVE AS TRD
SAVE AS TRD FACTOR uses the normalized measurement data to generate a
FACTOR
transducer factor with up to 625 points. The trace data are converted to a transducer
with unit dB after the transducer name has been entered. The number of points is
defined by SWEEP COUNT. The frequency points are allocated in equidistant steps
between start and stop frequency. The generated transducer factor can be further
adapted in the menu SETUP – TRANSDUCER. SAVE AS TRD FACTOR is only
available if normalization is switched on.
Remote command:
4.9.3

Reflection Measurement

Scalar reflection measurements can be carried out by means of a reflection-coeffi-
cient bridge.
Fig. 4.34 Test setup for reflection measurement
Operating Manual 1164.4556.12 - 06
External Generator Control – Option R&S FSP-B10
CORR:REC
CORR:TRAN:GEN <name>'
Instrument Functions
4.233