Manual Operation - Rohde & Schwarz R&S FSP Series Operating Manual

Spectrum analyzer
Table of Contents

Advertisement

R&S FSP
Two-port mixer:
P R E S E T
4.13.2

Manual Operation

The FREQ hardkey opens the menu for selecting the frequency and the frequency
range (details see
extended by the following functions for external mixing:
FREQ
EXTERNAL MIXER
4.266
.
.
S P E C T R U M A N A L Y Z E R
9 k H z . .. 3 G H z
F S P
L O O U T / IF IN
E x te rn a l
M ix e r
R F IN P U T
Connect the 'LO OUT / IF IN' output of the R&S FSP to the LO/IF port of the
external mixer. The nominal LO level is 15.5 dBm.
Because of the diplexer contained in the R&S FSP the IF signal can be tapped
from the line which is used to feed the LO signal to the mixer.
Feed the signal to be measured to the RF input of the external mixer.
"Frequency and Span Selection – FREQ Key" on page
!
EXT MIXER ON/OFF
SELECT BAND
CONV LOSS TABLE
ACCEPT BIAS
LO LEVEL
SIGNAL ID / AUTO ID /
AUTO ID THRESHOLD
LO/IF ports for external mixers - Option R&S FSP-B21
1 0 9 3 .4 4 9 5 . 0 3
M A D E I N G E R MA N Y
M A D E I N G E R MA N Y
L O / IF
R F
!
EDIT TABLE /
NEW TABLE
LOAD TABLE
DELETE TABLE
COPY TABLE
PAGE UP /
PAGE DOWN
Operating Manual 1164.4556.12 - 06
Instrument Functions
INSERT LINE
!
DELETE LINE
SAVE TABLE
PAGE UP /
PAGE DOWN
!
PAGE UP /
PAGE DOWN
4.9)