4.9.2.2
Normalization
NORMALIZE
The NORMALIZE softkey switches normalization on or off. The softkey is only avail-
able if the memory contains a correction trace.
It is possible to shift the relative reference point within the grid using the REF VALUE
POSITION softkey. Thus, the trace can be shifted from the top grid margin to the
middle of the grid:
Fig. 4.30 Normalized display
In the SPLIT SCREEN setting, the normalization is switched on in the current win-
dow. Different normalizations can be active in the two windows.
Normalization is aborted when the NETWORK mode is quit.
Remote command:
REF VALUE
The REF VALUE POSITION softkey (reference position) marks a reference position
POSITION
in the active window on which the normalization (difference formation with a refer-
ence curve) is performed.
When pressed for the first time, the softkey switches on the reference line and acti-
vates the input of its position. The line can be shifted within the grid limits.
The reference line is switched off by pressing the softkey again.
The function of the reference line is explained in the section
nism" on page
4.230
External Generator Control – Option R&S FSP-B10
CORR ON
4.234.
Instrument Functions
"Calibration Mecha-
Operating Manual 1164.4556.12 - 06