Markers and Delta Markers – MKR Key
The markers are used for marking points on traces, reading out measurement
results and for quickly selecting a display section. R&S FSP provides four markers
per display window. All markers can be used either as markers or delta markers.
The availability of marker functions depends on whether the measurement is per-
formed in the frequency, time or level domain.
The marker that you can move is defined in the following as the active marker.
marker
Fig. 4.7 Examples of marker display
Temporary markers are used in addition to the markers and delta markers to evalu-
ate the measurement results. They disappear when the associated function is deac-
tivated.
The measurement results of the active marker (also called marker values) are dis-
played in the marker field. The marker info field at the upper right of the display
shows the marker location (here, frequency), the level and the currently selected
trace [T1].
Operating Manual 1164.4556.12 - 06
active marker
1
3
2
delta marker
MARKER 1
[T1]
-27.5 dBm
123.4567 MHz
CAL:RES?
--
temporary marker
T1
Instrument Functions
Analyzer Mode
4.51