4.8 Tracking Generator – Option R&S FSP-B9
During normal operation (without a frequency offset), the tracking generator emits a
signal exactly at the input frequency of the R&S FSP.
For frequency-converting measurements it is possible to set a constant frequency
offset of ±150 MHz between the receive frequency of the R&S FSP and the output
signal of the tracking generator.
Moreover, an I/Q modulation or AM and FM modulation of the output signal can be
provided using two analog input signals.
The output power is level-controlled and can be set in 0.1 dB steps in a range from -
30 dBm to 0 dBm.
The tracking generator can be used in all operating modes. Acquisition of test setup
calibration values (SOURCE CAL) and normalization using these correction values
(NORMALIZE) is only possible in the NETWORK operating mode.
The RF characteristics of some DUTs is especially sensitive concerning the input
VSWR. In such cases insertion of 20 dB attenuation between the DUT and the
tracking generator output is highly recommended.
The tracking generator is activated by means of the NETWORK hotkey in the hotkey
bar at the bottom of the screen (for details refer to section
Bar" on page
4.8.1

Tracking Generator Settings

The NETWORK hotkey opens a menu for selecting the functions of the tracking
generator.
NETWORK
SOURCE ON/OFF
SOURCE POWER
POWER OFFSET
SOURCE CAL
FREQUENCY OFFSET
MODULATION
4.212
4.7.)
!
CAL TRANS
CAL REFL SHORT
CAL REFL OPEN
NORMALIZE
REF VALUE POSITION
REF VALUE
RECALL
SAVE AS TRD FACTOR
EXT AM
EXT FM
EXT I/Q
Instrument Functions
Tracking Generator – Option R&S FSP-B9
"Mode Selection – Hotkey
Operating Manual 1164.4556.12 - 06