Chapter 2 Mounting; Mount In Free-Stream Application (Flanged Fitting) - Emerson Micro Motion Installation Manual

Fork viscosity meters
Hide thumbs Also See for Micro Motion:
Table of Contents

Advertisement

2
Mounting
Topics covered in this chapter:
2.1
Mount in free-stream application (flanged
fitting)
Prerequisites
Free-stream (flanged) installations are recommended for processes with the following
conditions:
Flow
Viscosity
Temperature
Procedure
See
Figure 2-1
stream application.
Important
Always install the meter horizontally and oriented so that the gap between the tines is vertical. This
position helps to prevent the trapping of bubbles or solids on the meter – allowing the solids to sink
and the bubbles to rise. You can use the scribe mark on the spigot (located between the flange and
transmitter) as a reference for the tine orientation. Always orient the meter so the scribe mark is at
either the 12 o'clock or 6 o'clock position.
Installation Manual

Mount in free-stream application (flanged fitting)

Mount in free-stream application (weldolet fitting)
Mount with a T-piece (flanged fitting)
Mount with a flow-through chamber
Mount in an open tank (long-stem meter)
Mount in a closed tank (long-stem meter)
Attach the PFA ring and circlip
Rotate the electronics on the meter (optional)
Rotate the display on the transmitter (optional)
for information on installing the meter (with a flanged fitting) in a free-
0.3 to 0.5 m/s (at the meter)
0.5 to 12,500 cP
-50 °C to 200 °C (–58 °F to 392 °F)
-40 °C to 200 °C (-40 °F to 392 °F) in hazardous
areas
Mounting
9