3.3 Key Lock Function

3.4 Save and Recall Operations

– You can press [Shift]+[3] (Local) to switch the remote control to local
control. The mode modification will not affect the output parameters of
the power supply.
This function can prevent the power supply from the panel keys misoperation
during usage. Press the composite keys [Shift]+[2] (Lock) to lock front panel
keys and the lock symbol "*" is shown on the front panel display. All panel keys,
except the [On/Off] and [Shift]+[2] (Lock) keys, are locked. You can repress
[Shift]+[2] (Lock) to unlock the front panel.
The power supply can save up to 10 common parameters in nonvolatile memory
(No. 1 to No. 10) for user to recall conveniently. The saved settings include the
following parameters.
Category
Parameter
Voltage setting: Vs
Main interface
Current setting: Is
Upper limit of voltage: Vh, and lower limit of voltage: Vl
Upper limit of current: I+, and lower limit of current: I-
Upper limit of power: P+, and lower limit of power: P-
The output status of the instrument: [On/Off]
Config menu
CC/CV loop priority mode: Mode
CC/CV loop speed: Speed
Voltage/Current rise time: V-Rise Time/I-Rise Time
Voltage/Current fall time: V-Fall Time/I-Fall Time
On/Off switch delay: On Delay/Off Delay
Internal resistance of the power supply: Output Res
Protect menu
OCP/OVP/OPP/UCP/UVP switch status: On/Off
OCP/OVP/OPP/UCP/UVP limit setting: Level
Copyright © Itech Electronic Co., Ltd.
www.tehencom.com