Preparing The Detector For Transport; Transporting The Detector To A New Location - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

Follow these steps

7.13.1 Preparing the Detector for Transport

7.13.2 Transporting the Detector to a New Location

Charged Aerosol Detectors Corona Veo (RS) Operating Manual
1. Prepare the detector for transport. See the
2. The step depends as follows:
To transport the detector to a new location, follow the
instructions in
To ship the detector, follow the instructions in
page
204.
To prepare the detector for transport, follow these steps:
1. Perform a long-term shut down of the detector. See
page
156.
2. Turn off the detector with its main power switch and disconnect the
power cord.
3. Remove all cables and flow connections to other devices.
4. Remove the detector from the system stack. See
Removing the Detector from the System Stack, page
5. Remove the nebulizer. See
To transport the detector to a new location, follow these steps:
1. Observe the notes for handling and lifting the detector safely.
2. Transport the detector to the new location.
3. Install and set up the detector in the system stack. Follow the
instructions on mounting the system stack in
4. Set up the detector:
a) To connect the detector and to set up flow connections, follow
the instructions in the Installation chapter in this operating
manual.
7
section 7.13.2, page
203.
section 7.7, page
Maintenance and Service
section 7.13.1, page
section 7.13.3,
section 6.8.3,
section 7.12
200.
179.
section 5.5.1, page
Page 203
203.
67.