Digital I/O Terminals Pin Assignment - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

11
Appendix

11.2 Digital I/O Terminals Pin Assignment

Digital I/O Terminal A
Digital I/O Terminal B
Page 252
The detector is equipped with two I/O terminals that include 7
connection ports (4 TTL inputs and 3 relay outputs):
Pin
Signal Name
1
CC1+
2
CC1-
3
GND
4
CC2+
5
CC2-
6
GND
7
Divert+
8
Divert-
9
GND
Pin
Signal Name
1
Pump Off+
2
Pump Off-
3
GND
4
Gas Off+
5
Gas Off-
6
GND
7
Autozero+
8
Autozero-
9
GND
10
Start+
11
Start-
12
GND
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Signal Level
Remark
Relay output
Positive polarity
Relay output
Negative polarity
Ground
Relay output
Positive polarity
Relay output
Negative polarity
Ground
TTL input
Positive polarity
TTL input
Negative polarity
Ground
Signal Level
Remark
Relay output
Positive polarity
Relay output
Negative polarity
Ground
Relay output
Positive polarity
Relay output
Negative polarity
Ground
TTL input
Positive polarity
TTL input
Negative polarity
Ground
TTL input
Positive polarity
TTL input
Negative polarity
Ground