7.4.4
Cleaning or Decontaminating the Detector
Decontamination
Parts required
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Cleaning and decontamination must be performed by qualified
personnel wearing suitable personal protective equipment. Always
observe national and local regulations.
NOTICE Wipe up all liquids spilled onto the system immediately. If
surfaces are exposed for longer periods, these liquids can cause damage.
AVIS Essuyez immédiatement tout liquide renversé sur le système. Si les
surfaces sont exposées trop longtemps, ce liquide peut occasionner des
dégâts.
Decontamination is required, for example, when leakage or spillage has
occurred, or before service or transport of the detector. Use a suitable
cleaning detergent or disinfectant to ensure that the treatment renders
the detector safe to handle.
•
Suitable cleaning detergent (or disinfectant)
•
Purified water
•
Lint-free cloths or wipes
CAUTION—Explosive Gas Mixtures from Alcoholic Cleaning Detergents
Alcohol-containing cleaning detergents may form flammable and
explosive gas mixtures when exposed to air.
•
Use such cleaning detergents only when required and only in
adequately ventilated rooms.
•
Avoid open flames or exposure to excessive heat during the cleaning
process.
•
Wipe the cleaned components thoroughly dry after cleaning. Do not
operate the detector before it is completely dry.
7
•
Maintenance and Service