Troubleshooting
8.5

Operating Issues

8.5.1

Resolving Leakage

When
Parts required
Additional items required
Preparations
Follow these steps
Page 220
The leak sensor is wet. The leak sensor reports leakage.
Liquid may have collected in the leak tray under the flow connections.
Replacement part as required
Cloth or tissue
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in
Service, page
163.
1. Turn off the pump flow and then the gas flow:
a) Turn off the pump flow to the detector. Wait at least 5 minutes.
During the waiting period, the detector drains any remaining
mobile phase to waste.
b) Turn off the gas flow on the detector.
c) Wait until the system pressure is down to zero before you
continue.
d) Turn off the gas supply to the detector.
2. If you operate the detector from the chromatography data system,
disconnect the detector from the software.
3. Turn off the detector with the main power switch.
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
chapter 7 Maintenance and