Cleaning Stator And Rotor; Reassembling The Valve - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

7
Maintenance and Service
7.9.2

Cleaning Stator and Rotor

Items required
Preparations
Follow these steps
7.9.3

Reassembling the Valve

Parts required
Tools required
Preparations
Page 194
Deionized water
Microfiber cloth
Disassemble and inspect the valve. See
1. Dampen the microfiber cloth with deionized water.
2. Clean stator, rotor and the valve thoroughly with the microfiber
cloth. Take care that no surfaces are scratched while you are
cleaning the components.
After cleaning, it is not necessary to dry the rotor.
3. Reassemble the valve. See
If the stator is to be replaced: Stator
If the rotor is to be replaced: Rotor
Hexagon socket-head wrench, size 9/64"
1. Disassemble and inspect the valve. See
2. If the stator and/or rotor is to be replaced
Unpack the replacement stator and/or rotor.
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
section 7.9.1, page
section 7.9.3, page
194.
section 7.9.1, page
192.
192.