Updating The Detector Firmware - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

7
Maintenance and Service

7.11 Updating the Detector Firmware

When
Items required
Preparations
Page 198
The detector is shipped with the most recent firmware version.
To update the detector firmware, a chromatography data system, such
as the Chromeleon Chromatography Management System, is required.
The following update description refers to the Chromeleon software.
Updating the detector firmware might be required, for example, when a
new firmware version is released that adds functionality or solves
problems of a previous version.
Firmware version/Chromeleon Service Release as appropriate
TIP When a new firmware version is released, the new version will be
included in the next available Chromeleon Service Release. The new
firmware will not be transferred automatically to the device when you
install the Chromeleon Service Release.
Read the release notes provided with the firmware and/or
Chromeleon Service release.
Verify the following:
The detector is connected in Chromeleon.
All operations on the instrument (Chromeleon 7) or timebase
(Chromeleon 6.8) that includes the detector have been stopped.
The instrument or timebase is idle.
Check the currently installed firmware version in the detector:
On the Main Menu of the detector display
–or–
On the General tab page in the Chromeleon 7 Instrument
Configuration Manager or the Chromeleon 6.8 Server
Configuration program
Charged Aerosol Detectors Corona Veo (RS) Operating Manual