Signal Channels in Chromeleon

Page 156
Starting the new sequence
Select one of the following alternatives:
Start the sequence from the Sequence Control panel:
a) On the Sequence Control panel, click Load Sequence.
b) Select the sequence and click Open to add it to the batch list.
c) Click Start Batch.
Start the sequence from the Chromeleon 6.8 Browser:
a) In the Chromeleon 6.8 Browser, select the sequence.
b) On the Batch menu, click Edit > Add.
c) Select the sequence and click Open to add it to the batch list.
d) Click Start.
A Ready Check is performed. If the Ready Check passes and if the
timebase is not currently running another sequence, the sequence is
started.
The following signal channels are available for the detector in the
Chromeleon software:
CAD_1
The signal channel is available for data acquisition.
As a standard, the detector measures the current in pA.
Gas regulator pressure
The signal channel records and monitors the nebulizer gas pressure
during data acquisition.
As a standard, the detector measures the pressure in psi.
It is recommended to record the signal channel for troubleshooting
purposes.
For proper gas pressure regulation, or if baseline noise or poor
precision occurs, the information from the signal channel can
provide helpful information to identify and eliminate the source for
the problem.
Charged Aerosol Detectors Corona Veo (RS) Operating Manual