YOKOGAWA WT200 User Manual page 93

Digital power meter
Table of Contents

Advertisement

9.3 D/A Output (optional)
9-6
Relation between the output item and the D/A output voltage
• Frequency
D/A output
Approx. 7.5 V
5.0 V
2.5 V
0.5 V
0.2 V
4
10
• Integrated value
D/A output
Approx. 7.0 V
5.0 V
In case of 140% of
rated value input
0
• Other items
Displayed value
Output
140%
Approx. 7.0 V
100%
5.0 V
0%
0 V
–100%
–5.0 V
–140%
Approx. –7.0 V
However, for PF and deg, points in
the range from +5 to +7 V and from
–5 to –7 V are not output.
If there is an error, the output will
be about ±7.5 V.
For Vp and Ap, the output will be
±5 V when the value is three times
the range rating. In addition, output
will not be ±7.5 V when Vp and Ap
are over the range.
100
1 k
10 k 50 k
In case of rated
value input
D/A output
Approx. 7.5 V
Approx. 7.0 V
5.0 V
–140 –100
Displayed
value [Hz]
Time
t
o
t
:rated integration time
o
Displayed
value[%]
100 140
–5.0 V
Approx. –7.0 V
Approx. –7.5 V
IM 253421-01E