14.7 Communication-related Error Messages

Errors in communications commands (100 to 199)
Code Message
102
Syntax error
103
Invalid separator
104
Data type error
105
GET not allowed
108
Parameter not allowed
109
Missing parameter
111
Header separator error
112
Program mnemonic too long
113
Undefined header
114
Header suffix out of range
120
Numeric data error
123
Exponent too large
124
Too many digits
128
Numeric data not allowed
131
Invalid suffix
134
Suffix too long
138
Suffix not allowed
141
Invalid character data
144
Character data too long
148
Character data not allowed
150
String data error
151
Invalid string data
158
String data not allowed
161
Invalid block data
168
Block data not allowed
171
Invalid expression
178
Expression data not allowed
181
Invalid outside macro definition Does not conform to the macro definition specified in IEEE488.2.
IM 253421-01E
Error messages related to communications are given below.
When servicing is required, contact your nearest YOKOGAWA representative, as given
on the back cover of this manual.
Only error messages relating to the communication mode 488.2 are given here. For
other error messages, see chapter 13 and section 15.4, "Error Codes and Corrective
Actions."
Action
Incorrect syntax
Insert a comma between data items to separate them.
See pages 14-7 and 14-8 and enter data using the correct
data format.
GET is not supported as a response to an interface message.
Check the number of parameters.
Enter the required number of parameters.
Insert a space between the header and the data to separate them. 14-4
Check the mnemonic (character string consisting of letters and
numbers).
Check the header.
Check the header.
Mantissa must be entered before the numeric value in <NRf> format.
Use a smaller exponent in <NR3> format.
Limit the number of digits to 255 or less.
Enter in a format other than <NRf> format.
Check the units for <Voltage> and <Current>.
Check the units for <Voltage> and <Current>.
No units are allowed other than <Voltage> and <Current>.
Enter one of the character strings in {...|...|...}.
Check the character strings in {...|...|...}.
Enter in a format other than one of those in {...|...|...}.
<Character string> must be enclosed by double quotation marks
or single quotation marks.
<Character string> is too long or contains characters which cannot Section 14.3
be used.
Enter in a data format other than <Character string>.
<Block data> is not allowed.
<Block data> is not allowed.
Equation is not allowed.
Equation is not allowed.
Reference Page
or Section
Section 14.2,
14.3
14-4
14-7, 14-8
14-7,
Section 14.3
14-7,
Section14.3
Section 14.3
Section 14.3
Section 14.3
14-7
14-8,
Section 14.3
14-8,
Section 14.3
14-8,
Section 14.3
14-8
14-8
14-8
Section 14.3
Section 14.3
14
Section 14.3
14-8
Section 14.3
Section 14.3
Section 14.3
14-51

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents