10.1 GP-IB Interface Functions and Specifications
GP-IB Interface Specifications
10-2
• Electrical & mechanical specifications: conforms to IEEE St'd 488-1978
• Functional specifications: see the table below
• Code: ISO (ASCII) code
• Address setting: Address can be set in the range from 0 to 30 using the front panel
keys.
• Remote mode clear: remote mode can be cleared by pressing the LOCAL key on the
front panel. However, this is not possible when Local Lockout has been set by the
controller.
Function
source handshake
acceptor handshake
talker
listener
service request
remote local
parallel poll
device clear
device trigger
controller
The connectors used in this function have protective covers. When the covers are
removed or when using connectors, the voltage ratings across the measuring input
and the ground become as follows:
Voltage across A, ±(V and A side) input terminals and ground 400 Vrms max.
Voltage across V terminal and ground
Put the protective cover on the connector when this function is not used.
Subset name
Description
SH1
full source handshake capability
AH1
full acceptor handshake capability
T5
basic talker capability, serial polling, nontalker on
MLA (My Listen Address), talk-only capability
L4
Basic listener capability, nonlistener to MTA (My Talk
Address), no listen-only capability
SR1
full service request capability
RL1
full remote/local capability
PP0
no parallel polling capability
DC1
full device clear capability
DT1
full device trigger capability
C0
no controller function
CAUTION
600 Vrms max.
IM 253421-01E