MEASure:HARMonics:ITEM:{SYNChronize|
<harmonic analysis function>}
Function
Sets the communication output item of
harmonic analysis ON/OFF, queries the current
setting.
Syntax
MEASure:HARMonics:ITEM:{SYNChronize|
<harmonic analysis
function>} {<Boolean>}
MEASure:HARMonics:ITEM:{SYNChronize|
<harmonic analysis function>}?
SYNChronize=PLL source
<harmonic analysis function>={VTHD|V|
VCON|ATHD|A|ACON|PF|W|WCON|VDEG|ADEG}
Example
MEASURE:HARMONICS:ITEM:VTHD ON
MEASURE:HARMONICS:ITEM:VTHD?→:MEASURE:
HARMONICS:ITEM:VTHD 1
Description
The selection
SYNChronize
frequency of the PLL source.
You can query the PLL source input by the
command
HARMonics:SYNChronize?
MEASure:HARMonics:VALue?
Function
Queries harmonic analysis data set by
commands other
than"
MEASure:HARMonics:ITEM
Syntax
MEASure:HARMonics:VALue?
Example
MEASURE:HARMONICS:VALUE?→60.00E+00,
12.01E+00,49.98E+00,49.62E+00,0.03E+00,
5.50E+00,.....
Description
• The renewal of harmonic analysis data output
here occurs when bit0 (UPD) of the condition
register (see section 14.4) changes from high
to low. For details, see section 14.2.6,
"Synchronization with the Controller."
• For the output format of harmonic analysis
data, see page 14-30.
MEASure:NORMal?
Function
Queries all settings related to normal
measured/computed data.
Syntax
MEASure:NORMal?
Example
MEASURE:NORMAL?→:MEASURE:NORMAL:ITEM:
V:ELEMENT1 1;:MEASURE:NORMAL:ITEM:A:
ELEMENT1 1;:MEASURE:NORMAL:ITEM:W:
ELEMENT1 1;:MEASURE:NORMAL:ITEM:VA:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:VAR:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:PF:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:DEGREE:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:VHZ:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:AHZ:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:WH:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:WHP:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:WHM:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:AH:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:AHP:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:AHM:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:VPK:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:APK:
ELEMENT1 0;:MEASURE:NORMAL:ITEM:TIME 0;
MATH 0
IM 253421-01E
MEASure[:NORMal]:ITEM?
Function
Syntax
Example
MEASure[:NORMal]:ITEM:PRESet
Function
Syntax
Example
Description
is for outputting the
MEASure[:NORMal]:ITEM:{TIME|MATH}
Function
".
Syntax
Example
MEASure[:NORMal]:ITEM:<normal
measurement function>?
Function
Syntax
Example
14.3 Commands
Queries all settings related to the
communication output items of normal
measured/computed data.
MEASure[:NORMal]:ITEM?
MEASURE:NORMAL:ITEM?→
(Results are the same
as for
)
MEASure:NORMal?
Sets the ON/OFF pattern for all communication
outputs of the normal measurement function.
MEASure[:NORMal]:ITEM:PRESet {NORMal|
INTEGrate|CLEar}
MEASURE:NORMAL:ITEM:PRESET NORMAL
The following three patterns can be selected.
NORMal: V/A/W→ON, others→OFF
INTEGrate: W/WH/AH/TIME→ON,
others→OFF
CLEar: all items→OFF
Sets the communication output of {elapsed time
of integration|MATH} ON/OFF, queries about
the current setting.
MEASure[:NORMal]:ITEM:{TIME|
MATH} {<Boolean>}
MEASure[:NORMal]:ITEM:{TIME|MATH}?
MEASURE:NORMAL:ITEM:TIME OFF
MEASURE:NORMAL:ITEM:TIME?→:MEASURE:
NORMAL:ITEM:TIME 0
Queries communication output settings of the
normal measurement function.
MEASure[:NORMal]:ITEM:<normal
measurement function>?
<normal measurement
function>={V|A|W|VA|VAR|PF|DEGRee|VHZ|AHZ|
WH|WHP|WHM|AH|AHP|AHM|VPK|APK}
MEASURE:NORMAL:ITEM:V?→:MEASURE:NORMAL:
ITEM:V:ELEMENT1 1