6.4
Precautions Regarding Use of Integrator
Function
Relation between Integration Hold and the START/STOP key
Relation between Integration Reset and the START/STOP key
Integrated
value
Elapsed time
of integration
START
IM 253421-01E
When the HOLD key is pressed, the display and communication output of the integrated
results is being held while integration continues. The relation between this hold function
and the START/STOP key is as follows.
• Even when starting integration while the hold function is on, the display and
communication output will remain unchanged. Only canceling the hold function or
activating a trigger (pressing the SHIFT key followed by the HOLD (TRIG) key) will
result in displaying or outputting the integrated results of the time of cancellation.
ON
HOLD
OFF
Displayed
value
(Dotted line shows integrated value)
Elapsed time
of integration
START
• Even when stopping integration while the hold function is on, the displayed integrated
value will remain unchanged. However, as soon the hold function is turned off or a
trigger is activated, the integrated results of the time when integration was stopped will
be displayed or output.
ON
HOLD
OFF
ON
TRIG
Displayed
value
(Dotted line shows integrated value)
Elapsed time
of integration
START
The relation between integration reset and the start/stop key is as follows.
Interrupt
Interrupt
Restart
Start
STOP
START
STOP
STOP RESET
ON
ON
STOP
RESET
Reset
Restart
Integration timer preset time
RESET
START
6
Auto stop
Reset
RESET
6-9