Recall Group - YOKOGAWA WT200 User Manual

Digital power meter
Table of Contents

Advertisement

14.3 Commands

14.3.10 RECall Group

The commands in the RECall group are used to make settings relating to, and inquires about recalling data. This
allows you to make the same settings and inquiries as can be set using the lower menus of [MEMORY]-"rECAL" or
[MEMORY]-"PnLrC".
:RECall
:
RECall?
Function
Queries all the settings relating to recalling
data.
Syntax
RECall?
Example
RECALL?→:RECALL:STATE 0;INTERVAL 0,0,0
RECall:INTerval
Function
Sets the recalling interval/queries the current
setting.
Syntax
RECall:INTerval
{<NRf>,<NRf>,<NRf>|<String>}
RECall:INTerval?
{<NRf>,<NRf>,<NRf>}=0,0,0 to 99,59,59
{<String>}=HH:MM:SS
SS seconds
Example
RECALL:INTERVAL 0,0,0
RECALL:INTERVAL "00:00:00"
RECALL:INTERVAL?→:RECALL:INTERVAL 0,0,0
Description
Even when the interval has been set to 0,0,0,
the interval becomes 250ms in case of normal
measurement and 1s in case of harmonic
analysis.
RECall:PANel
Function
Retrieves the setting parameters file.
Syntax
RECall:PANel {<NRf>}
{<NRf>}=1 to 4 : file number
Example
RECALL:PANEL 1
RECall[:STATe]
Function
Turns recalling ON/OFF, queries the current
setting.
Syntax
RECall[:STATe] {<Boolean>}
RECall:STATe?
Example
RECALL:STATE ON
RECALL:STATE?→:RECALL:STATE 1
14-32
;
STAT e
<Space>
<NRf>
?
INTer val
<Space>
<NRf>
<Character String>
?
PANel
<Space>
<NRf>
?
HH hour MM minutes
OFF
ON
,
<NRf>
,
<NRf>
IM 253421-01E

Hide quick links:

Advertisement

Table of Contents
loading