13.1 Commands
RO/RO?
Sets the recall function ON/OFF or
inquires about the current setting.
Syntax
RO m<terminator>
"m" indicates recall ON or OFF.
m=
0:
recall OFF
1:
recall ON
Query
RO?<terminator>
Example
RO1
Description
Parameter error 12 will occur if "m" is set to an
illegal value.
RR/RR?
Sets the recall interval or queries the
current setting.
Syntax
RR m1,m2,m3<terminator>
"m1" indicates the hour
0 ≤ m1 ≤ 99
"m2" indicates the minutes
0 ≤ m2 ≤ 59
"m3" indicates the seconds
0 ≤ m3 ≤ 59
Query
RR?<terminator>
Example
RR0,0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• When the recall interval is set to 0 hrs, 0 min, 0
sec, the interval will be 250 msec during normal
measurement and 1s during harmonic analysis.
• While recalling or storing is in progress,
execution error 19 will occur.
RT/RT?
Sets the rated integration time when
integrated values are to be output as an
analog signal or queries the current setting.
Syntax
RT m1,m2,m3<terminator>
"m1" indicates hour, and must be set
within the following range.
0 ≤ m1 ≤ 10000
"m2" indicates minute, and must be set
within the following range.
0 ≤ m2 ≤ 59
"m3" indicates second, and must be set
within the following range.
0 ≤ m3 ≤ 59
Query
RT?<terminator>
Example
RT1,0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• The maximum selectable time is 10000 (hours).
RV/RV?
Sets voltage range or queries the
current setting.
Syntax
RV m<terminator>
"m" indicates voltage range.
m=
3:
15V range
4:
30 V range
5:
60 V range
7:
150 V range
8:
300 V range
9:
600 V range
13-10
Query
RV?<terminator>
Example
RV9
Description • Parameter error 12 will occur if an illegal value
is set.
• Changing of the voltage range is not allowed
while integration is in progress; execution
error 13 will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
SA/SA?
Sets the external sensor scaling
constant or queries the current setting.
Syntax
When CM0 is set:
SA n<terminator>
When CM1 is set:
SA m,n<terminator>
"m" indicates element.
m=
1:
Element 1
"n" indicates external sensor scaling
constant.
0.001 ≤ n ≤ 9999.
Query
When CM0 is set:
SA?<terminator>
When CM1 is set:
SAm?<terminator>
Example
When CM0 is set:
SA50.00
When CM1 is set:
SA1,50.00
Description • Parameter error 12 will occur if "m" is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SC/SC?
Determines whether or not to use the
scaling function or queries the current
setting.
Syntax
SC m<terminator>
"m" indicates whether scaling is ON or
OFF.
m=
0:
OFF
1:
ON
Query
SC?<terminator>
Example
SC1
Description • Parameter error 12 will occur if "m" is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SL
Recalls setting parameters from a
selected file.
Syntax
SL m<terminator>
"m" indicates file no., and must be set
within the following range.
1 ≤ m ≤ 4
Description • Parameter error 12 will occur if "m" is set to an
illegal value.
• It is not possible to recall communications-
related information (communication mode,
address etc.) using this command.
• While recalling or storing is in progress,
execution error 19 will occur.
IM 253421-01E