5.1 Displaying Voltage, Current, and Active Power
Explanation
5-2
Continuous Maximum Allowable Input
• Voltage
Up to a peak value of 1.5 kV or RMS value of 1.0 kV, whichever is less.
• Current
• 20 A to 0.5 A range
Up to a peak value of 100 A or RMS value of 25 A, whichever is less.
• 200 mA to 5 mA range
Up to a peak value of 30 A or RMS value of 20 A, whichever is less.
• External sensor input
Up to five times the measurement range.
Maximum Reading of the Display and Units
• Maximum reading : for voltage, current and power, each 99999 (when the number
of displayed digits is 5)
• Units: V (voltage), A (current), W (power)
• Prefix: m, k, or M
Selecting the Display Function
The following selections are available.
• V: voltage will be displayed
• A: current will be displayed
• W: power will be displayed
Note
The measurement results may differ even on a same signal if the measurement mode is
changed. For details related to the measurement mode, see section 4.1.
The displayed values of V (voltage), A (current), and W (active power), while the MAX hold
function (see section 4.8) is enabled, will be the maximum values (MAX) that are held. The
values for D/A output, output to external plotter and printer, and communication output are
also set to the maximum values (MAX) that are held.