IM 253421-01E
5. Stopping Integration
Press the STOP key. The START indicator will extinguish and the STOP indicator will
light. The displayed values will be held.
STOP
6. Resetting Integration
Press the RESET key. The STOP indicator will extinguish and the values on display
A will be reset to 0.00.00.
RESET
Maximum Reading of the Display and Units
Maximum reading
• Integrated value: 999999 (–99999 in case of minus display)
• Elapsed time of integration: 10000
• Units: Wh (power integration: watt hour value), Ah(current integration: ampere hour
value)
• Prefix: m, k, or M
Selecting the Display Function
The following selections are available.
• Wh: displays both the positive and negative watt hour values
• Wh±: displays the positive and negative watt hour values separately
• Ah: displays the total ampere hour values
• Ah±: displays the total ampere hour values, or the positive and negative ampere
hour values separataly
: The average active power during integration is displayed, if the function is
specified in the computation settings (see section 4.11).
Update Hold Function
Although the held values will not be updated, integration continues. When hold is
being cancelled, the integration results (values and time) corresponding to the point of
cancellation, will be displayed.
For details regarding the relation with the START/STOP key, refer to the following
section.
Integration Reset
Resetting will result in returning the integration results to the status before integration
started.
Pressing the RESET key is useful after integration has been stopped.
For details regarding the relation with the START/STOP key, refer to the following
section.
Display in case of Integration Over
When the maximum integration value has been reached (999999 MWh/MAh or
–99999 MWh/MAh), integration will stop and that result will be held on the display.
When the maximum integration time has been reached (up to 10000 hrs), integration
will stop and that result will be held on the display.
6.3 Displaying Integrated Values