3.2

Installing the Instrument

Installation Conditions
Installation Position
3-2
The instrument must be installed in a place where the following conditions are met.
Ambient temperature and humidity
• Ambient temperature: 5 to 40˚C
• Ambient humidity: 20 to 80% RH (no condensation)
Horizontal position
The instrument must be installed horizontally. A non-horizontal or inclining position can
impede proper measurement of the instrument.
Well-ventilated location
Vent holes are provided on the top and bottom of the instrument. To prevent rise in internal
temperature, do not block these vent holes.
In case you removed the feet for rack-mounting the instrument, make sure to keep a space of
at least 20 mm as not to block the vent holes.
Never install the instrument in any of the following places
• In direct sunlight or near heat sources;
• Near noise sources such as high voltage equipment or power lines ;
• Where an excessive amount of soot, steam, dust or corrosive gases is present;
• Where the level of mechanical vibration is high;
• Near magnetic field sources;
• In an unstable place.
Note
To ensure high measurement accuracy, the instrument should only be used under the following
conditions.
Ambient temperature: 23 ± 5˚C
Ambient humidity: 30 to 75% RH (no condensation)
When using the instrument in the temperature ranges of 5 to 18 or 28 to 40˚C, add the temperature
coefficient to the accuracy as specified in chapter 16 "Specifications".
If the ambient humidity of the installation site is 30% or below, use an anti-static mat to prevent
generation of static electricity.
Internal condensation may occur if the instrument is moved to another place where both ambient
temperature and humidity are higher, or if the room temperature changes rapidly. In such cases
acclimatize the instrument to the new environment for at least one hour before starting operation.
Desktop
As shown below, place the instrument on a flat even surface. When using the handle for
installation, check that the handle is in one of the fixed positions. To change the fixed position of
the handle, pull the handle approximately 2 to 3 mm outward along the rotational axis and slowly
move the handle.
Fixed positions of the handle
(We recommend the positions 1, 3, 5, or 8. When
using no 4, don't put any weight on the instrument.)
7
8
6
5
1
4
2
3
3
Turning axis
Turn the stands after
pulling them approx.
2-3 mm on both sides.
1
IM 253421-01E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents