YOKOGAWA WT200 User Manual page 141

Digital power meter
Table of Contents

Advertisement

13.1 Commands
DA/DA?
Sets the function for display A or
queries the current setting.
Syntax
DA m<terminator>
"m" indicates one of the following
functions.
• During normal measurement
m=
1:voltage (V)
2:current (A)
3:power (W)
4:reactive power (var)
5:apparent power (VA)
15:Integration time (TIME)
• During harmonic analysis
m=
1:
Each relative harmonic content
of 1st to 50 (or 30) th order of
voltage (V)
2:
Each relative harmonic content
of 1st to 50 (or 30) th order of
current (A)
3:
Each relative harmonic content
of 1st to 50 (or 30) th order of
active power (W)
28: harmonic analysis order (order)
Query
DA?<terminator>
Example
DA1
Description
Parameter error 12 will occur if "m" is set to an
illegal value.
DB/DB?
Sets the function for display B or
queries the current setting.
Syntax
DB m<terminator>
"m" indicates one of the following
functions.
• During normal measurement
m=
1:voltage (V)
2:current (A)
3:power (W)
6:power factor (PF)
11:phase angle (deg)
• During harmonic analysis
m=
1:
Analysis value of each component
of voltage (V)
2:
Analysis value of each component
of current (A)
3:
Analysis value of each component
of active power (W)
6:
power factor (PF)
16: harmonic distortion factor of
voltage (V THD)
17: harmonic distortion factor of
current (A THD)
19: Relative harmonic content of
each voltage component (V %)
20: Relative harmonic content of
each current component (A %)
21: Relative harmonic content of each
active power component (W %)
22: Phase angle between each voltage
of the 2nd to 50 (or 30) th
order and the fundamental (1st
order) voltage.
23: Phase angle between each current
of the 2nd to 50 (or 30) th
order and the fundamental (1st
order) current.
Query
DB?<terminator>
Example
DB1
Description
Parameter error 12 will occur if "m" is set to an
illegal value.
13-2
DC/DC?
Sets the function for display C/inquires
about the current setting
Syntax
DC m<terminator>
"m" indicates one of the following
functions.
• During normal measurement
m=
1:voltage (V)
2:current (A)
3:power (W)
7:Input voltage frequency (V Hz)
8:Input current frequency (A Hz)
9:watt hour (Wh)
10:ampere hour (Ah)
12:Peak voltage value (Vpk)*
13:Peak current value (Apk)*
14:Computation result (MATH)*
24:positive watt hour (Wh+)
25:negative watt hour (Wh–)
26:positive ampere hour (Ah+)
27:negative ampere hour (Ah–)
• During harmonic analysis
m=
1:
Rms value of the 1st to 50 (or
30) th order of voltage (V)
2:
Rms value of the 1st to 50 (or
30) th order of current (A)
3:
Rms value of the 1st to 50 (or
30) th order of active power (W)
7:
Input voltage frequency (V Hz)
8:
Input current frequency (A Hz)
Query
DC?<terminator>
Example
DC1
Description
Parameter error 12 will occur if "m" is set to an
illegal value.
DF/DF?
Sets the computation method for
harmonic distortion (THD) or queries the
current setting.
Syntax
DF m<terminator>
"m" indicates the computation method for
harmonic distortion (see to section 7.2)
m=
0:IEC
1:CSA
Query
DF?<terminator>
Example
DF0
Description • Parameter error 12 will occur if "m" is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
DL/DL?
Sets the terminator for communication
output data or queries the current setting.
Syntax
DL<terminator>
"m" indicates terminator
GP-IB
m=
0:
CR+LF+EOI
1:
LF
2:
EOI
Query
DL?<terminator>
Example
DL0
RS-232-C
CR+LF
LF
CR
IM 253421-01E