Type L in °C
Units
decimal
900.0
9000
:
:
-200.0
-2000
< -200.0
-32768
Measuring range for thermocouple type N
Table 13-24 SIMATIC S7 format: Measuring range type N in °C and °F
Type N in °C
Units
decimal
> 1550.0
32767
1550.0
15500
:
:
1300.1
13001
1300.0
13000
:
:
-270.0
-2700
< -270.0
-32768
Measuring range for thermocouple types R, S
Table 13-25 SIMATIC S7 format: Measuring range type R, S in °C and °F
Type R, S in
Units
°C
decimal
> 2019.0
32767
2019.0
20190
:
:
1769.1
17691
1769.0
17690
:
:
-50.0
-500
-50.1
-510
:
:
-170.0
-1700
< -170.0
-32768
ET 200iSP
Operating Instructions, 11/2017, A5E00247483-07
Type L in °F
hexadec‐
imal
2328
1652.0
H
:
:
F830
-328.0
H
8000
< -328.0
H
Type N in °F
hexadec‐
imal
7FFF
> 2822.0
H
3C8C
2822.0
H
:
:
32C9
2372.1
H
32C8
2372.0
H
:
:
F574
-454.0
H
8000
-32768
H
Type R, S in
°F
hexa‐
decimal
decimal
7FFF
> 3276.6
32767
H
4EDE
3276.6
32766
H
:
:
:
451B
3216.3
32163
H
451A
3216.2
32162
H
:
:
:
FE0C
-58.0
-580
H
FE0B
-58.1
-581
H
:
:
:
F95C
-274.0
-2740
H
8000
< -274.0
-32768
H
Units
decimal
hexa‐
decimal
16520
4088
Nominal range
H
:
:
-3280
F330
H
-32768
80000
Underflow
H
Units
decimal
hexa‐
decimal
32767
7FFF
Overflow
H
28220
6E3C
Overrange
H
:
:
23721
5CA9
H
23720
5CA8
Nominal range
H
:
:
-4540
EE44
H
8000
<EE44
Underflow
H
H
Units
hexa‐
decimal
7FFF
Overflow
H
7FFE
Overrange
H
:
7DA3
H
7DA2
Nominal range
H
:
FDBC
H
FDBB
Underrange
H
:
F54C
H
8000
Underflow
H
Analog electronic modules
13.8 Representation of analog values
Range
Range
Range
299

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents