Step 5. Enter Your Elevation
Step 5. Enter Your Elevation
In order for your Keysight 86120B to accurately measure wavelengths and meet its
published specifications, you must enter the elevation where you will be performing
your measurements.
1 Press the Setup key.
2 Press the MORE softkey.
3 Press the CAL softkey.
4 Press ELEV.
5 Use the
and
meter steps from 0 m to 5000 m.
The elevation value selected with the softkeys must be within 250 meters of the ac-
tual elevation.
6 Press RETURN to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by
using the following equation:
20
softkeys to enter the elevation in meters. Entries jump in 500
------------ -
m
=
3.281
ft