Attribute Summary
Description
Syntax
Attribute Summary
Query Response
Programming Commands
SENSe Subsystem
Argument
Description
AIR
Selects wavelength values in standard air.
VACuum
Selects wavelength values in a vacuum.
Preset State: VAC
*RST State: VAC
SCPI Compliance: instrument specific
Standard air is defined to have the following characteristics:
Barometric pressure: 760 torr
Temperature: 15°C
Relative humidity: 0%
CORRection:OFFSet[:MAGNitude]
Enters an offset for amplitude values.
:SENSe:CORRection:OFFSet:MAGNitude{?| {<real> | MINimum | MAXimum}}
<real> is the logarithmic units in dB.
Constant
Description
−40.0 dB
MINimum
MAXimum
40.0 dB
Preset State: 0.0
*RST State: 0.0
SCPI Compliance: standard
The query form returns the current offset setting as shown in the following example:
+5.00000000E+000
DATA?
Queries the time domain samples of the input laser line.
204