Query Response
Example
Programming Commands
Common Commands
The service request enable register is cleared when the instrument is turned on. The
*RST and *CLS commands do not change the register. The *SRE? query returns the
value of the service request enable register.
Table 5-17. Service Request Enable Register
Bit
Bit Weight
a
7
128
6
64
5
32
4
16
3
8
2
4
1
2
0
1
a. High enables the status byte register bit.
<integer> from 0 to 63 or from 128 to 191.
OUTPUT 720;"*SRE 32"
In this example, the command enables ESB (event summary) bit 5 in the status byte
register to generate a service request.