Syntax
Attribute Summary
Description
WLIMit:STOP:WNUMber
Sets the stopping wavenumber for the wavelength limit range.
:CALCulate2:WLIMit:STOP:WNUMber {?|{<real> | MINimum | MAXimum }}
<real> is a wavenumber value that is within the following limits:
Constant
Description
MINimum
start wavelength limit
MAXimum
14286 cm
Non-sequential command
Preset State: 8.333335E5 m
*RST State: 8.333335E5 m
SCPI Compliance: instrument specific
This command sets the stopping range for the wavelength limit. The default units for
the <real> parameter are inverse meters. The stop wavenumber value must be less
than or equal to the start wavenumber value or the stop wavenumber will be clipped
to the start wavenumber and a "Data out of range" error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to "Always force the Keysight 86120B to wait for non-sequential com-
mands" on page 92
for more information.
(700 nm)
-1
-1
-1