Keysight Technologies 86120B User Manual page 153

Multi-wavelength meter
Table of Contents

Advertisement

<resolution>
Constants
Examples
Query Response
MAXimum
0.01 resolution (fast update)
MINimum
0.001 resolution (normal)
DEFault
Current resolution
:CONF:ARR:POW:FREQ DEF MIN
:FETC:ARR:POW:FREQ? DEF MAX
:READ:ARR:POW:FREQ?
:MEAS:ARR:POW:FREQ?
:CONF:SCAL:POW:FREQ 230.8THZ, MAX
:FETC:SCAL:POW:FREQ? 230.8THZ, MIN
:READ:SCAL:POW:FREQ? 230.8THZ
:MEAS:SCAL:POW:FREQ? 230.8THZ
The following line is an example of a returned string when
:MEAS:SCAL:POW:FREQ? MAX is sent:
+1.94055176E+014
If six laser lines are located and :MEAS:ARR:POW:FREQ? is sent, the following
string is an example of the returned data. Notice that the first returned number indi-
cates the number of laser-line values returned in the query.
6,+1.94055100E+014,+1.93854100E+014,+1.93653000E+014,+1.93452000E+014,+1.9325
0900E+014,+1.93050000E+014
Programming Commands
Measurement Instructions
151