Keysight Technologies 86120B User Manual page 150

Multi-wavelength meter
Table of Contents

Advertisement

Syntax
Description
<expected_value>
Constants
Programming Commands
Measurement Instructions
MEASure{:ARRay | [:SCALar]} :POWer?
Returns amplitude values.
:POWer? [<expected_value>[,<resolution>]]
Used With
<expected_value>
SCALar
optional
ARRay
ignored
When used with a :SCALar command, a single value is returned. The display is
placed in the single-wavelength mode, and the marker is placed on the signal having
a power level that is closest to the <expected_value> parameter.
When used with an :ARRay command, an array of amplitudes is returned. The dis-
play is placed in the list-by-power mode.
Returned values are in the current power units. Wavelength units are not affected.
CONFigure command
When this function is used with the CONFigure command, the query question
mark character "?" must not be included in the string. However, the FETCh,
READ, and MEASure command are queries and require the question mark.
Refer to the examples for this command.
MAXimum
Displays the highest power signal
MINimum
Displays the lowest power signal
DEFault
Displays the signal at the current marker position
148
<resolution>
ignored
ignored

Advertisement

Table of Contents
loading

Table of Contents