Description
Example
Syntax
Attribute Summary
Description
Example
Changes in the state of a condition register bit causes the associated OPERation Sta-
tus or QUEStionable Status register bit to be set. This command allows you to select
a negative bit transition to trigger an event to be recognized. A negative transition is
defined to occur whenever the selected bit changes states from a 1 to a 0. You can
enter any value from 0 to 65535.
When queried, the largest value that can be returned is 32767. This is because the
most-significant register bit cannot be set true.
OUTPUT 720;":STATUS:OPER:NTRansition 16"
{OPERation | QUEStionable}:PTRansition
Selects bits in the event register which can be set by positive transitions of the corre-
sponding bits in the condition register.
:STATus:OPERation:PTRansition{?| <integer>}
<integer> an integer from 0 to 65535.
Preset State: none
*RST State: none
SCPI Compliance: standard
Changes in the state of a condition register bit causes the associated OPERation Sta-
tus or QUEStionable Status event register bit to be set. This command allows you to
select a positive bit transition to trigger an event to be recognized. A positive transi-
tion is defined to occur whenever the selected bit changes states from a 0 to a 1. You
can enter any value from 0 to 65535.
When queried, the largest value that can be returned is 32767. This is because the
most-significant register bit cannot be set true.
OUTPUT 720;":STATUS:OPER:PTRansition 16"