Syntax
Attribute Summary
Description
Query Response
Programming Commands
SENSe Subsystem
CORRection:DEVice
Selects the wavelength measurement algorithm. This command applies to
Keysight 86120B instruments with firmware version number 2.0. When first turned
on, the instrument briefly displays the firmware version. Instruments with a firm-
ware version number less than 2.0 do not have this feature.
:SENSe:CORRection:[DEVice]{?| {NARRow | BROad}}
Constant
Description
NARRow
Selects wavelength measurements for narrowband devices such
as DFB lasers and modes of FP lasers.
BROad
Selects wavelength measurements for broadband devices such
as optical filters and LEDs.
Non-sequential command
Preset State:
ow
NARR
*RST sets this value to NARRow.
SCPI Compliance: instrument specific
The narrow bandwidth algorithm, used for measuring lasers, determines the wave-
length based upon the peak.
The broad bandwidth algorithm, used for LEDs, filters, and chirped lasers, deter-
mines the wavelength based upon the center-of-mass of the power spectrum. The
peak excursion function is used to determine the value of the integration limits. Care
must be taken to ensure that the integration limits are above any noise. This is espe-
cially true when measuring devices with sloping noise floors, like an EDFA ampli-
fier. For more information on peak excursion,
Instrument specifications apply when the device is set to
do not apply in
ad mode.
BRO
The query form returns the previously selected device.
ow
NARR
202
refer to "PEXCursion" on page
ow. Specifications
NARR
164.