Syntax
Attribute Summary
Description
:CALCulate2:WLIMit[:STATe]{?| {ON | OFF | 1 | 0}}
Non-sequential command
Preset State: on
*RST State: on
SCPI Compliance: instrument specific
When this function is on, the Keysight 86120B has an input range from the WLIMit
STARt to the WLIMit STOP. When this function is off, the instrument displays
peaks over the full wavelength range. If you want to measure signals over a nar-
rower wavelength range, set this function on to avoid identifying spurious second
harmonic peaks.
Whenever the Keysight 86120B receives this command, it reprocesses the data and
performs a new peak search.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to "Always force the Keysight 86120B to wait for non-sequential com-
mands" on page 92
for more information.
Programming Commands
CALCulate2 Subsystem
167