Syntax
Attribute Summary
Description
DRIFt:DIFFerence[:STATe]
Sets the drift calculation to subtract the minimum values measured from the maxi-
mum values measured.
:CALCulate3:DRIFt:DIFFerence[:STATe]{?| {ON | OFF | 1 | 0}}
Preset State: off
*RST State: off
SCPI Compliance: instrument specific
Use the CALC3:DRIF:PRES command to turn off all the drift states before turning
on this state. The CALC3:DATA? query returns the maximum power and frequency
minus the minimum power and frequency.
Note
Only one STATe command can be turned on at any one time. Attempting to turn
more than one state on at a time results in a "–221 Settings Conflict" error.
Refer to "Measure delta, drift and signal-to-noise" on page 94
information on selecting measurements.
Programming Commands
CALCulate3 Subsystem
for additional
185