SNR:AUTO
Selects the reference frequency value for measuring noise in the signal-to-noise cal-
culation.
:CALCulate3:SNR:AUTO{?| {ON | OFF | 1 | 0}}
Constant
Description
ON
Selects internally generated reference frequency.
OFF
Selects user-entered reference frequency.
Preset State: on
*RST State: on
SCPI Compliance: instrument specific
The command argument allows you to select either an internally generated or a user-
entered frequency reference for measuring the noise. To enter a value to use as the
reference, use the SNR:REFerence:FREQuency, SNR:REFerence:WAVelength, and
SNR:REFerence:WNUMber commands.