SA (Spectrum Analyzer) Mode
SA Mode Settings

IF Output

The FieldFox IF (intermediate frequency) can be output for external signal
processing. The signal is available from the IF Out connector on the FieldFox
right-side panel. See the connector in
The IF Out signal, centered at approximately 33.75 MHz, is simply a
down-converted version of the RF Input signal that is present at the tuned
frequency.
While outputting an IF signal, the FieldFox stops performing the InstAlign
process. Therefore, the amplitude of traces on the FieldFox screen is NOT
accurate and
in
Learn more about the IF Output frequencies and bandwidths in the
Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
How to select IF Output
The IF Output signal is useful only in Zero Span. Learn more about Zero Span in
"Zero Span Measurements" on page
At least one sweep must be made in order to tune the FieldFox.
— Press BW 2.
— Then
— Then
— Then choose from the following:
Keysight N9927-90001 User's Guide
—Auto
The FieldFox selects the most efficient sweep type based on the
Res BW. When Res BW is set to 200 kHz and below, FFT sweep type
is selected. Above 200 kHz, Step sweep type is being performed.
—FFT
The FieldFox uses FFT sweep type regardless of the Res BW.
—Step
The FieldFox uses Step sweep type regardless of the Res BW.
This is useful for capturing impulsive noise that may be generated
from nearby electrical equipment.
Meas UNCAL
"Individual Alignments" on page
Advanced
IF Output Use
No output from the IF Out connector.
—OFF
—Narrow
The IF output signal has approximately 5 MHz bandwidth.
—Wide
The IF output signal has approximately 25 MHz bandwidth.
"Right Side Panel" on page
appears on the screen to indicate this. Learn more
180.